Hirst GM09

£0.00

The GM09 is a compact gaussmeter offering general testing capabilities with hall probes.

  • Compact Size

  • Touchscreen

  • Data Logging

  • Rechargeable Li-ion Battery

The GM09 is a compact gaussmeter offering general testing capabilities with hall probes.

  • Compact Size

  • Touchscreen

  • Data Logging

  • Rechargeable Li-ion Battery

Available Services:Supply Hire Calibration Repair


The Hirst GM09 Gaussmeter offers versatile measurement capabilities with support for transverse, axial, and 3D Hall probes, ensuring compatibility with a wide range of magnetic field sensing applications. It maintains backward compatibility with existing Hirst probes, allowing seamless integration with previously acquired equipment. Its compact design facilitates easy handheld operation, complemented by a responsive touchscreen interface for intuitive control. The device is powered by a rechargeable lithium-ion battery, conveniently charged through a USB-C port, delivering reliable performance in a portable form factor.

Specification

Measurement rangesAUTO, ±3 mT, ±30 mT, ±300 mT, ±3.000 T
RepeatabilityBetter than 0.5%
AccuracyBetter than ±1% — calibrated to standards traceable to the NPL (National Physical Laboratory, UK)
Resolution2 µT, 0.02 Gauss (Oe), 1.6 A/m
Colours and materialsAnodised aluminium with plastic end caps
ConnectivityUSB-C port for charging & data retrieval (mass storage .CSV time-stamped files)
Dimensions / Weight139 × 88 × 49 mm / 390 g
Maximum field sensed±3 T (dependent on probe)
Measurement modesDC (with Min & Max), AC-Peak (with AC-Peak-Max), AC-RMS (with AC-RMS-Max)
Number of sensorsSingle channel, 1-axis sensor probe
PowerLi-ion 3500 mAh, USB-C chargeable — >4 hours continuous (with sleep & auto power-off)
Standard cable length1.5 m
TimeReal time clock with date/time — battery backup >3 years
UnitsTesla, Gauss, Oersted (Oe), A/m — (1 mT = 10 Gauss = 0.7967 kA/m)
Frequency RangeStatic field (DC) and AC 10 Hz – 6 kHz
Display Rate3 readings per second
Display3.5″ multi-colour touch screen, 320 × 480 px, backlight
Temperature CoefficientBetter than ±0.1% of reading / °C (including probe)
Optimum Operating Temperature20 °C – 40 °C
Optimum Storage Temperature0 °C – 50 °C