Olympus Epoch 600 2016-12-06T11:57:33+00:00

Olympus Epoch 600 Ultrasonic Flaw Detector

This is a Olympus Epoch 600 ultrasonic flaw detector, available for hire / rental from VARE
Hire Now
  • Calibration

  • Repair

  • Supply

  • Accessories

Brochure
Operator Manual
Gates 2 Independent Gates
Display Delay -59 mm to 13401 mm (-2.323 in. to 527.6 in.) @ longitudinal velocity in steel
Zero 0 µs to 750 µs
Velocity 635 m/s to 15240 m/s (0.0250 in./µs to 0.6000 in./µs)
Range 3.36 mm to 13,405 mm (0.132 in. to 527 in.) @ 5,900 m/s (0.2320 in./μ)
Units MM, Inches, Microseconds (µs)
Test Modes Pulse Echo, Dual, or Through Transmission
Display Full VGA (640 x 480 pixels) transflective color LCD, 60 Hz update rate. 117 mm x 89 mm, 146 mm (4.62 in. x 3.49 in., 5.76 in.)
Pulser Tuneable Square Wave
PRF 10 Hz to 2000 Hz in 10 Hz increments
Energy settings 100 V, 200 V, 300 V or 400 V
Pulse width Adjustable from 25 ns to 5,000 ns (0.1 MHz) with PerfectSquare™ technology
Damping 50, 100, 200, 400 Ω
Gain 0 to 110 dB
Receiver bandwidth 0.2 MHz to 26.5 MHz at -3 dB
Digital filter settings Eight digital filter sets standard (0.2-10 MHz, 2.0-21.5 MHz, 8.0-26.5 MHz, 0.5-4 MHz, 0.2-1.2 MHz, 1.5-8.5 MHz, 5-15 MHz, DC-10 MHz)
Rectification Full-wave, Positive Half-wave, Negative Half-wave, RF
Horizontal linearity ± 0.5% FSW
Reject 0 to 80% FSH with Visual Warning
Battery Single lithium-ion rechargeable standard, 12 h to 13 h (lithium-ion), 3 h (alkaline)
Storage 50,000 IDs on board, removable 2 GB MicroSD card (standard)
Auto Cal Velocity, Zero Offset Straight Beam (First Backwall or Echo-to-Echo) Angle Beam (Sound path or Depth)
Languages English, Spanish, French, German, Japanese, Chinese, Portuguese
Weight 1.68 kg (3.72 lb), including lithium-ion battery
Dimensions 236 mm x 167 mm x 70 mm (9.3 in. x 6.57 in. x 2.76 in.)
Probe Connectors Lemo (BNC Adaptors available upon request)